1 | Red and Infrared Emission Tube |
2 | Red and Infrared Receipt Tube |
•Fingernail polish or false fingernails. | 1 |
Display Type | LED (Light Emitting Diode) | ||||||
SpO2 | Measurement | 70-99% | |||||
range | |||||||
Accuracy | 80%-99%: ±2% | ||||||
70%-79%: ±3% | |||||||
≤69% : no definition | |||||||
Pulse Rate | Measurement | 30-235 BPM | |||||
range | |||||||
Accuracy | 30~99 BPM: ±2 BPM | ||||||
100~235 BPM: ±2% | |||||||
Pulse Intensity | Bargraph Indicator | ||||||
Power Requirement | Two AAA alkaline Batteries | ||||||
Power Consumption | <40 mA | ||||||
Low Power Indicator | |||||||
Battery Life | ~ 30 hours of continuous operation | ||||||
Dimension | 2.20″ ~ 2.44″ x 1.26″ ~ 1.50″ x 1.34″ ~ 1.50″ | ||||||
(L x W x H) | (56 mm ~ 62 mm x 32 mm ~ 38 mm x 34 mm ~ 38 mm) | ||||||
Weight | 1.59 oz ~ 2.12 oz (0.10 lb ~ 0.13 lb) (45 g ~ 60 g) | ||||||
including two AAA batteries | |||||||
Environmental | Temperature | Operation | 41°F ~ 104°F | ||||
Requirements | (5°C ~ 40°C) | ||||||
Storage | 68°F ~ 131°F | ||||||
(20°C ~ 55°C) | |||||||
Humidity | Operation | ≤80% RH | |||||
(non-condensing) | |||||||
Storage | ≤93% RH | ||||||
Interference | Device works normally when mixed noise produced | ||||||
Resistance Capacity | by BIO-TEK INDEX Pulse Oximeter tester According to initial rumors, this smartphone will be powered Snapdragon 625, but some call Galaxy A6+ will provided 450 Snapdragon.FeaturesEarlier, the news of the Galaxy A6 was first known from the known WiFi certification last month. Samsung galaxy tab 7 plus user manual. For its performance, this smartphone has 4 GB of RAM with an internal memory of 32 GB.ProcessorHowever, it is not yet known how it performs for the chipset used. These devices will be supported chipsets Exynos 7870, RAM 3 GB, and 32 GB of internal memory.CameraWhile Galaxy A6+ will carry the same resolution with Samsung Galaxy A6, but has a larger screen size, which is 6 inches. Samsung Galaxy A6 2018 User Manual PDFSamsung Galaxy A6 2018 manual – After the information about the whereabouts of alleged smartphone Samsung Galaxy A6 2018 appeared last month, now turn the smartphone specs leaked on the internet.DesignQuoted from GSM Arena, Sunday (8/4/2018), Galaxy A6 2018 referred will have a screen size 5.6 inch Infinity Display with a resolution of was about 1,280 x 2,220. | ||||||
againstAmbient Light |
EmissionTest | Compliance | Electromagnetic environment – guidance | ||||
RF emission | Group 1 | The Pulse Oximeter uses RF energy only for its | ||||
CISPR 11 | internal function. Therefore, its RF emissions | |||||
are very low and are not likely to cause any | ||||||
interference in nearby electronic equipment. | ||||||
RF emission | Class B | The Pulse Oximeter is suitable for use | ||||
CISPR 11 | in all establishments, including domestic | |||||
establishments and those directly connected to | ||||||
the public low-voltage power supply network that | ||||||
supplies buildings used for domestic purposes. | ||||||
TROUBLESHOOTING | ||||||
Problem | Possible reason | Solution | ||||
SpO | 2 | or PR | 1. User's finger is incorrectly | 1. Reinsert user's finger | ||
inserted | 2. Attempt several times to | |||||
cannot be | ||||||
displayed | 2. User’s Oxyhemoglobin | obtain a reading; If sure that | ||||
normally | value is too low to be | no problem exists, obtain | ||||
measured | further clinical examination Sony cyber shot 8.1. | |||||
SpO | 2 | or PR | 1. Finger may not be | 1. Reinsert finger | ||
inserted deeply enough | 2. Ask user to remain still | |||||
display is | ||||||
unstable | 2. Finger trembling or user | |||||
moving |
Troubleshooting continued | ||||||||||
Problem | Possible reason | Solution | ||||||||
Oximeter | 1. | Battery power may be | 1. | Replace batteries | ||||||
cannot be | inadequate or batteries | 2. | Reinstall batteries | |||||||
powered on | may not be installed | 3. | Contact GF distributor | |||||||
2. | Batteries may be installed | |||||||||
incorrectly | ||||||||||
3. | Oximeter may be damaged | |||||||||
Indicator | 1. | Device automatically | 1. | Normal | ||||||
lamps are | powers off when no signal | 2. | Replace the batteries | |||||||
suddenly | is detected for longer than | |||||||||
off | 8 seconds | |||||||||
2. | Batteries too weak to | |||||||||
power device | ||||||||||
“Error3” or | 1. | Low power | 1. | Replace batteries | ||||||
“Error4” | 2. | Receiving tube and/or | 2. | Contact GF distributor | ||||||
displays | connector may be | 3. | Contact GF distributor | |||||||
shielded or damaged | 4. | Contact GF distributor | ||||||||
3. | Mechanical misplace for | |||||||||
receive-emission tube | ||||||||||
4. Amp circuit malfunction | ||||||||||
“Error7” | 1. | Low power | 1. | Replace batteries | ||||||
displays | 2. | Emission diode damaged | 2. | Contact GF distributor | ||||||
3. | Current control circuit | 3. | Contact GF distributor | |||||||
malfunction | ||||||||||
SYMBOL DEFINITIONS | ||||||||||
Symbol | Definition | Symbol | Definition | |||||||
Type BF applied part | Low power indicator | |||||||||
Attention, consult | No SpO2 Alarm | |||||||||
accompanying documents | ||||||||||
SpO2 | Oxygen saturation | | Power switch | |||||||
Heart rate (BPM) | SN | Serial Number | ||||||||
BPM |
© 2010 GF Health Products, Inc., Atlanta GA 30360, 770-368-4700 | JB02017-INS-LAB-RevA10 |